Model 7800
ECG
Lead Selection:Trigger Lead: LI, LII, LIII, and AUTO - menu selectable.
Second Lead: LI, LII and LIII – menu selectable.
Patient Cable: 4-lead patient cable with 6-Pin AAMI standard connector.
Isolation: Isolated from ground related circuits by >4 kV rms, 5.5 kV peak
Electrode Impedance Measurement
Technique: 10 Hz ac signal < 10 uA rmsRange: 200kO per lead
Accuracy: ±3% ±1kO
Recommended Electrode: 10% Chloride sponge type (Ivy P/N: 590436)
Cardiotach
Range: 10 to 350 BPM (Pediatric / Neonate)10 to 300 BPM (Adult)
Accuracy: ±1% ±1 BPM
Resolution: 1 BPM
Sensitivity: 300 µV peak
Tall T Wave Rejection: Rejects T waves = 1.2 * R-wave
Pacer Pulse Rejection
Width: 0.1 to 2 ms at ±2 to ±700 mVAlarms
High Rate: 15 to 250 BPM in 5 BPM incrementsLow Rate: 10 to 245 BPM in 5 BPM increments
Asystole: R to R interval >6 seconds
Lead Off: Detached lead or offset potential >0.5 V
Test Mode
Internal: ECG 1 mV/100 ms referred to input @ 70 BPMSimulator: ECG waveform amplitude: 1mV
Simulator Range: 10 – 250 BPM.
Simulator Rate: Variable rate in steps of 30, 60, 90, 120, 150, 180, 210 and 240 BPM. Manually adjustable in increments of 1 BPM.
Display
Active Matrix TFT Color Touch Screen LCD (640x480)Screen Size: 17.09cm x 12.82cm, 21.36cm (8.4") diagonal
Mechanical
Size: Height: 8.72in. (22.14cm)Width: 9.25in. (23.50cm)
Depth: 6.10in. (15.49cm)
Weight: 5.6 lbs. (2.54kg)
Recorder
Optional FeatureWriting Method: Direct Thermal
Synchronized Output (Trigger)
Test input signal at ECG leads: Conditions: ½ sine wave, 60ms width, 1mV amplitude, 1 pulse/secondOutput Trigger Delay: < 10 ms
R to R Trigger Accuracy: ±75 µs typical @ 1 mV input
Pulse width: 100ms
Pulse amplitude: 0 to +5V
Output Impedance: <100 O
Operating Environment
Temperature Range: 15°C to 40°CRelative Humidity: 0-90% non-condensing
Altitude: -100 meters to +3,000 meters
Atmospheric Pressure: 500-1060 mbar
Protection against ingress of fluids: IPX1 – Protection against vertically falling drops of water
Storage Environment
Temperature Range: -20°C to 49°CRelative Humidity: 0-100%
Altitude: -100 meters to +14,000 meters
Power Requirements
Voltage Input: 100-120V~; 200-230V~Line Frequency: 50/60 Hz
Fuses Type and Rating: T.5A, 250V (Metric 5x20 mm)
Maximum ac Power
Consumption: 45 VA
Power Recovery: Automatic, if power is restored within 30 seconds
Communications
Ethernet 10Base-T, IEEE 802.3 (2 Channels)RS-232
Data Storage
USB Memory StickRegulatory
Unit meets or exceeds the specifications for:- AAMI Cardiac Monitor Standard EC-13
- UL60601-1
- CAN/CSA C22.2 No 601.1-M90
- CDN MDR (CMDCAS)
- IEC 60601-2-27
- MDD.93/42/EEC
- CE 0413
- ISO 13485:2003
- FDA/CGMP