Model 7810

Technical Specifications

ECG

Lead Selection:
Trigger Lead: LI, LII, LIII, and AUTO - menu selectable.
Second Lead: LI, LII and LIII – menu selectable.
Patient Cable: 4-lead patient cable with 6-Pin AAMI standard connector.
Isolation: Isolated from ground related circuits by >4 kV rms, 5.5 kV peak

Frequency Response

LCD Display and Recorder
Filtered 1.5 to 40 Hz or 3.0 to 25 Hz
Unfiltered 0.67 to 100 Hz
ECG Output
Unfiltered 0.67 to 100 Hz

Electrode Impedance Measurement

Technique: 10 Hz ac signal < 10 uA rms
Range: 200kO per lead
Accuracy: ±3% ±1kO
Recommended Electrode: 10% Chloride sponge type (Ivy P/N: 590436)

Cardiotach

Range: 10 to 350 BPM (Pediatric / Neonate)
10 to 300 BPM (Adult)
Accuracy: ±1% ±1 BPM
Resolution: 1 BPM
Sensitivity: 300 µV peak
Tall T Wave Rejection: Rejects T waves = 1.2 * R-wave

Pacer Pulse Rejection

Width: 0.1 to 2 ms at ±2 to ±700 mV

Alarms

High Rate: 15 to 250 BPM in 5 BPM increments
Low Rate: 10 to 245 BPM in 5 BPM increments
Asystole: R to R interval >6 seconds
Lead Off: Detached lead or offset potential >0.5 V

Test Mode

Internal: ECG 1 mV/100 ms referred to input @ 70 BPM
Simulator: ECG waveform amplitude: 1mV
Simulator Range: 10 – 250 BPM.
Simulator Rate: Variable rate in steps of 30, 60, 90, 120, 150, 180, 210 and 240 BPM. Manually adjustable in increments of 1 BPM.

Recorder

Feature Optional
Writing Method Direct Thermal

Display

Active Matrix TFT Color Touch Screen LCD (640x480)
Screen Size: 17.09cm x 12.82cm, 21.36cm (8.4") diagonal

Mechanical

Size: Height: 8.72in. (22.14cm)
Width: 9.25in. (23.50cm)
Depth: 6.10in. (15.49cm)
Weight: 5.6 lbs. (2.54kg)

Respiratory Performance

Pressure Range: 0 to 2.5 psi @ 760 mm Hg
Digitization Rate: 1KHz
Waveform transmission Rate: 100Hz (configurable)
Digitization Resolution: 16 bit
Bandwidth: .05 to 3Hz
Baseline restoration: < 10 seconds
Gain: AGC or Manual

Respiratory Patient Connection

Pneumatic connection: Colder Products #PMC1602
Hose to pillow connection: Lure Lock
Pillow strap: 2” wide Velcro Strap

Synchronized Output (Trigger)

Test input signal at ECG leads: Conditions: ½ sine wave, 60ms width, 1mV amplitude, 1 pulse/second
Output Trigger Delay: < 10 ms
R to R Trigger Accuracy: ±75 µs typical @ 1 mV input
Pulse width: 100ms
Pulse amplitude: 0 to +5V
Output Impedance: <100 O

Operating Environment

Temperature Range: 5°C to 40°C
Relative Humidity: 0-90% non-condensing
Altitude: -100 meters to +3,600 meters
Atmospheric Pressure: 500-1060 mbar
Protection against ingress of fluids: IPX1 – Protection against vertically falling drops of water

Storage Environment

Temperature Range: -40°C to 70°C
Relative Humidity: 5% to 95%
Altitude: -100 meters to +14,000 meters

Power Requirements

Voltage Input: 100-120V~; 200-230V~
Line Frequency: 50/60 Hz
Fuses Type and Rating: T.5A, 250V (Metric 5x20 mm)
Maximum ac Power
Consumption: 45 VA
Power Recovery: Automatic, if power is restored within 30 seconds

Communications

Ethernet RJ-45 10Base-T, IEEE 802.3 (2 Channels)

Data Storage

USB Flash Drive, minimum capacity 1GB
ECG storage: 200 most recent events

Regulatory

Unit meets or exceeds the specifications for:
  • AAMI Cardiac Monitor Standard EC-13
  • UL60601-1
  • CAN/CSA C22.2 No 601.1-M90
  • CDN MDR (CMDCAS)
  • IEC 60601-2-27
  • MDD.93/42/EEC
  • CE 0413
  • ISO 13485:2003
  • FDA/CGMP