Model 7810
ECG
Lead Selection:Trigger Lead: LI, LII, LIII, and AUTO - menu selectable.
Second Lead: LI, LII and LIII – menu selectable.
Patient Cable: 4-lead patient cable with 6-Pin AAMI standard connector.
Isolation: Isolated from ground related circuits by >4 kV rms, 5.5 kV peak
Frequency Response
LCD Display and RecorderFiltered 1.5 to 40 Hz or 3.0 to 25 Hz
Unfiltered 0.67 to 100 Hz
ECG Output
Unfiltered 0.67 to 100 Hz
Electrode Impedance Measurement
Technique: 10 Hz ac signal < 10 uA rmsRange: 200kO per lead
Accuracy: ±3% ±1kO
Recommended Electrode: 10% Chloride sponge type (Ivy P/N: 590436)
Cardiotach
Range: 10 to 350 BPM (Pediatric / Neonate)10 to 300 BPM (Adult)
Accuracy: ±1% ±1 BPM
Resolution: 1 BPM
Sensitivity: 300 µV peak
Tall T Wave Rejection: Rejects T waves = 1.2 * R-wave
Pacer Pulse Rejection
Width: 0.1 to 2 ms at ±2 to ±700 mVAlarms
High Rate: 15 to 250 BPM in 5 BPM incrementsLow Rate: 10 to 245 BPM in 5 BPM increments
Asystole: R to R interval >6 seconds
Lead Off: Detached lead or offset potential >0.5 V
Test Mode
Internal: ECG 1 mV/100 ms referred to input @ 70 BPMSimulator: ECG waveform amplitude: 1mV
Simulator Range: 10 – 250 BPM.
Simulator Rate: Variable rate in steps of 30, 60, 90, 120, 150, 180, 210 and 240 BPM. Manually adjustable in increments of 1 BPM.
Recorder
Feature OptionalWriting Method Direct Thermal
Display
Active Matrix TFT Color Touch Screen LCD (640x480)Screen Size: 17.09cm x 12.82cm, 21.36cm (8.4") diagonal
Mechanical
Size: Height: 8.72in. (22.14cm)Width: 9.25in. (23.50cm)
Depth: 6.10in. (15.49cm)
Weight: 5.6 lbs. (2.54kg)
Respiratory Performance
Pressure Range: 0 to 2.5 psi @ 760 mm HgDigitization Rate: 1KHz
Waveform transmission Rate: 100Hz (configurable)
Digitization Resolution: 16 bit
Bandwidth: .05 to 3Hz
Baseline restoration: < 10 seconds
Gain: AGC or Manual
Respiratory Patient Connection
Pneumatic connection: Colder Products #PMC1602Hose to pillow connection: Lure Lock
Pillow strap: 2” wide Velcro Strap
Synchronized Output (Trigger)
Test input signal at ECG leads: Conditions: ½ sine wave, 60ms width, 1mV amplitude, 1 pulse/secondOutput Trigger Delay: < 10 ms
R to R Trigger Accuracy: ±75 µs typical @ 1 mV input
Pulse width: 100ms
Pulse amplitude: 0 to +5V
Output Impedance: <100 O
Operating Environment
Temperature Range: 5°C to 40°CRelative Humidity: 0-90% non-condensing
Altitude: -100 meters to +3,600 meters
Atmospheric Pressure: 500-1060 mbar
Protection against ingress of fluids: IPX1 – Protection against vertically falling drops of water
Storage Environment
Temperature Range: -40°C to 70°CRelative Humidity: 5% to 95%
Altitude: -100 meters to +14,000 meters
Power Requirements
Voltage Input: 100-120V~; 200-230V~Line Frequency: 50/60 Hz
Fuses Type and Rating: T.5A, 250V (Metric 5x20 mm)
Maximum ac Power
Consumption: 45 VA
Power Recovery: Automatic, if power is restored within 30 seconds
Communications
Ethernet RJ-45 10Base-T, IEEE 802.3 (2 Channels)Data Storage
USB Flash Drive, minimum capacity 1GBECG storage: 200 most recent events
Regulatory
Unit meets or exceeds the specifications for:- AAMI Cardiac Monitor Standard EC-13
- UL60601-1
- CAN/CSA C22.2 No 601.1-M90
- CDN MDR (CMDCAS)
- IEC 60601-2-27
- MDD.93/42/EEC
- CE 0413
- ISO 13485:2003
- FDA/CGMP